DBCG-3351/1151LT flange liquid level transmitter

Click for original picture
Fob Price: Discuss personally 
Payment Terms: delivery withindays since payment
Location: China Shanghai
  Wu Qi Jian    Contact Now 询价
Not exactly what you want? Post a quick Buying Request !
 
 
Product Detail

Origin:China Shanghai

Product detail: DBCG-3351/1151LT flange liquid level transmitter ] 

 

 


1. Main features

Complete in varieties and specifications, high precision, good stability

Two-wire system (four-wire system in special situations)

Sensitive elements, plug-in circuit boards, components can be interchanged with imported 1151 products of the same kind.

The range and zero position can be adjusted continuously externally

Positive transfer up to 500%, negative transfer up to 600%;

Adjustable damping;        

3-digit LCD liquid crystal display and pointer indicator show the measurement results

2. Function parameters

Object: liquid, gas and steam
Measuring range: see selection table

Output signal: 4 ~ 20mADC ( four-wire system220V, AC power supply, 0 ~ 10mADC output)
Power supply: 12 ~ 45VDC, usually 24VDC
Load characteristics: related to the power supply, with load capacity at a certain power supply voltage, the relation between load impedance RL and power source voltage VS is RL<50 (VS-12)
Display instrument: pointer type linear indication 0 ~ 100% scale and square root indication 0-10 scale, and the 3-digit liquid crystal display (0 
 100% linear), 3 LED (0-100% linear or range mode)

Range and zero: externally, continuously adjustable
Positive and negative transfer: after the zero has gone through positive transfer or negative transfer, the absolute value of the upper limit and lower limit of the range and measuring range can not exceed 100% of the upper limit of the measuring range; maximum positive transfer amount should be 500% of the minimum 
adjusting rangemaximum negative transfer amount for 500% of the minimum adjusting rangemaximum positive and negative transfer amount of the flow transmitter should be 10% of calibration range.

Temperature range: general transmitter-25C ~ +70C.
Static pressure and overload pressure: the differential pressure transmitter can bear maximum static working pressure, and the transmitter will not be damaged.
Volume variation: <0.
16cm3

Damping time: generally adjustable between 0.2S to 1.67S
Start-up time: 2S no need preheating

Models and specifications (Selection Guide)

Table 1

 

D    DBCG-3351/1151LT flange liquid level transmitterMax working pressure2.5MPa

 

 

 

 

 

     

 

 

 

 

 

 

 

 

Code

Range(kpa)

 

4

0-6.237.4

 

5

0-31.1186.8

 

6

0-117690

 

 

 

Code

Output

 

E

Ordinary

 

S

Intelligent

 

 

 

 Code

Nominal diameter ()

Inserting tube length()

Material for isolation diaphragm on the high voltage side

 

A0

80

Flat

316LSST

 

A2

80

50

316LSST

 

A4

80

100

316LSST

 

A6

80

150

316LSST

 

B0

100

Flat

316LSST

 

B2

100

50

316LSST

 

B4

100

100

316LSST

 

B6

100

150

316LSST

 

C0

80

Flat

Hastelloy C -276

 

C2

80

50

Hastelloy C -276

 

C4

80

100

Hastelloy C -276

 

C6

80

150

Hastelloy C -276

 

D0

100

Flat

Hastelloy C -276

 

D2

100

50

Hastelloy C -276

 

D4

100

100

Hastelloy C -276

 

D6

100

150

Hastelloy C -276

 

E0

80

Flat

Tantalum

 

F0

100

Flat

Tantalum

 

 

 

Code

Installing flange

 

A

31501b

 

B

41501b

 

C

33001b

 

D

43001b

 

 

 

Code

Structure material

 

Flange joint

Exhaust/drain valve

Isolation Diaphragm

Filling liquid

 

22

316Stainless steel

316Stainless steel

316Stainless steel

Silicone

Oil

 

23

316Stainless steel

316Stainless steel

HastelloyC alloy-276

 

24

316Stainless steel

316Stainless steel

Monel

 

25

316Stainless steel

316Stainless steel

Tantalum

 

33

HastelloyC alloy

HastelloyC alloy

HastelloyC alloy-276

 

35

HastelloyC alloy

HastelloyC alloy

Tantalum

 

 

 

Code

Accessory and additional functions

 

M1

0—100% linear meter

 

M3

3LCD digital display

 

D1

Pressure chamber bleed valve located at the upper

 

D2

Pressure chamber bleed valve located at the lower

 

None

NPT taper pipe female thread connecting block

 

C1

NPT pressure leading joint, back welding φ14 pressure leading pipe

 

C2

M20×1.5 T-shaped male thread connecting block

 

C21

M20×1.5 T-shaped male thread connecting block with additional T-shaped joint

 

d

Flameproof dBT4

 

i

Intrinsically safe ibCT6

 

 

 

 

D   DBCG-3351/1151LT 5 E     A6    A    22     M1              ←Selection example

 

                                       

 

 
more..Other products from this supplier

[ ProductsSearch ]  [ [ Print ]  [ Close ]


Copyright © 2004 ChinaAseanTrade.com All rights reserved
中国-东盟自由贸易网版权所有@2004 沪lCP证05013507号

沪公网安备 31010102002066号